DocumentCode
1021396
Title
Calibration of network analyser measurements with leakage errors
Author
Heuermann, Holger ; Schiek, B.
Author_Institution
Inst. fur Hochfrequenztech., Ruhr-Univ., Bochum
Volume
30
Issue
1
fYear
1994
fDate
1/6/1994 12:00:00 AM
Firstpage
52
Lastpage
53
Abstract
Useful and accurate MMIC wafer probe measurements can only be performed when the network analyser is calibrated at the device under test (DUT) interface and all leakage errors and coupling terms are considered in the error model. To measure the parameters of an MMIC on a wafer, the calibration procedure must be based on the full model. Two simple calibration procedures for determining error correction parameters for the full model are presented. Both 15 term procedures are based on closed solutions and only need three known calibration standards, so that the accuracies are excellent, as proved by measurements
Keywords
MMIC; calibration; error correction; integrated circuit testing; measurement errors; microwave measurement; network analysers; 15 term procedures; MMIC wafer probe measurements; calibration procedures; closed solutions; error correction parameters; error model; leakage errors; network analyser measurements;