  • Abstract
    An AC bridge readout system has been developed that greatly improves the low-frequency stability of bolometric detectors. The readout can be implemented with a simple circuit appropriate for use in space applications. A matched pair of detectors was used in the readout to achieve system noise within a factor of two of the fundamental noise limit of the detectors at frequencies as low as 10 mHz. The low-frequency stability of the readout system allows slower, more sensitive detectors to be used in many applications, and it facilitates observing strategies that are well suited to space-borne observations
  • Keywords
    bolometers; bridge circuits; frequency stability; 10 mHz; AC bridge readout system; bolometric detectors; fundamental noise limit; low frequency stability; matched pair; space applications; space-borne observations; system noise; Frequency; Heat sinks; Infrared detectors; Infrared heating; Low-frequency noise; Optical noise; Radiation detectors; Semiconductor device noise; Stability; Telescopes;