• DocumentCode
    1045085
  • Title

    A study of data loss and mispositioning due to pileup in 2-D detectors in PET

  • Author

    Germano, G. ; Hoffman, E.J.

  • Author_Institution
    Div. of Nucl. Med. & Biophys., California Univ., Los Angeles, CA, USA
  • Volume
    37
  • Issue
    2
  • fYear
    1990
  • fDate
    4/1/1990 12:00:00 AM
  • Firstpage
    671
  • Lastpage
    675
  • Abstract
    The effect of event pileup was investigated with a PET (positron emission tomography) system which employs a 2-D detector-array system. Event pileup caused losses of resolution of 9%, 19.8%, and 32% of the FWHM (full width at half maximum) for 10, 20, and 30 mCi of positron emitter in the field of view (FOV), respectively. 9.4%, 18.7%, and 28.1% of the data from the outer planes of the module were incorrectly placed in the inner planes for 5, 10, and 15 mCi of the positron emitter in the FOV, respectively. A simple model of the process was formulated and found to be adequate to describe the major features of event pileup for a number of source configurations and a wide range of activities
  • Keywords
    biomedical equipment; computerised tomography; radioisotope scanning and imaging; 2D detectors; PET; data loss; detector-array system; inner planes; mispositioning; outer planes; pileup; positron emission tomography; positron emitter; resolution; source configurations; Detectors; Energy resolution; Event detection; Image resolution; Loss measurement; Nuclear medicine; Photomultipliers; Photonic crystals; Positron emission tomography; Signal processing;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.106696
  • Filename
    106696