• DocumentCode
    1060031
  • Title

    Temperature coefficient variations in heat treated Evanohm and their effect on the transient behavior of the NML 1 Ω resistors

  • Author

    Pritchard, Brian J. ; Small, Greig W.

  • Author_Institution
    Nat. Meas. Lab., CSIRO, Lindfield, NSW, Australia
  • Volume
    42
  • Issue
    2
  • fYear
    1993
  • fDate
    4/1/1993 12:00:00 AM
  • Firstpage
    557
  • Lastpage
    561
  • Abstract
    The NML 1-Ω Evanohm resistors have a transient thermal response that is a function of the resistor element and the local thermal environment. The temperature coefficient of the Evanohm coil varies systematically from top to bottom as well as in an arbitrary pattern turn by turn. In the presence of a temperature gradient the resistance value may be different from that when the temperature is uniform. This temperature gradient should be less than 30 mK if the change resistance is to be less than 0.01 p.p.m
  • Keywords
    electric resistance measurement; heat treatment; measurement standards; resistors; transient response; 1 ohm; Evanohm coil; NML; fabrication; heat treated Evanohm; local thermal environment; temperature coefficient; temperature gradient; transient behavior; transient thermal response; Coils; Electrical resistance measurement; Heat treatment; Petroleum; Q measurement; Resistance heating; Resistors; Temperature distribution; Temperature sensors; Thermal resistance;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.278625
  • Filename
    278625