DocumentCode
1060031
Title
Temperature coefficient variations in heat treated Evanohm and their effect on the transient behavior of the NML 1 Ω resistors
Author
Pritchard, Brian J. ; Small, Greig W.
Author_Institution
Nat. Meas. Lab., CSIRO, Lindfield, NSW, Australia
Volume
42
Issue
2
fYear
1993
fDate
4/1/1993 12:00:00 AM
Firstpage
557
Lastpage
561
Abstract
The NML 1-Ω Evanohm resistors have a transient thermal response that is a function of the resistor element and the local thermal environment. The temperature coefficient of the Evanohm coil varies systematically from top to bottom as well as in an arbitrary pattern turn by turn. In the presence of a temperature gradient the resistance value may be different from that when the temperature is uniform. This temperature gradient should be less than 30 mK if the change resistance is to be less than 0.01 p.p.m
Keywords
electric resistance measurement; heat treatment; measurement standards; resistors; transient response; 1 ohm; Evanohm coil; NML; fabrication; heat treated Evanohm; local thermal environment; temperature coefficient; temperature gradient; transient behavior; transient thermal response; Coils; Electrical resistance measurement; Heat treatment; Petroleum; Q measurement; Resistance heating; Resistors; Temperature distribution; Temperature sensors; Thermal resistance;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.278625
Filename
278625