• DocumentCode
    1060097
  • Title

    A 10 V calibrating system using an automated voltage divider and a 1 V Josephson array

  • Author

    Sakamoto, Yasuhiko ; Endo, Tadashi ; Sakuraba, Toshiaki

  • Author_Institution
    Electrotech. Lab., Ibaraki, Japan
  • Volume
    42
  • Issue
    2
  • fYear
    1993
  • fDate
    4/1/1993 12:00:00 AM
  • Firstpage
    583
  • Lastpage
    587
  • Abstract
    A 10-V calibrating system is developed by combining an automated voltage divider and a 1-V Josephson junction array voltage standard. The system can calibrate the 10 V output of a Zener voltage reference with an uncertainty of 0.02 ppm
  • Keywords
    Josephson effect; calibration; measurement standards; voltage dividers; voltage measurement; 10 V; Josephson array; Zener voltage reference; automated voltage divider; Calibration; Circuits; Electrical resistance measurement; Helium; Josephson junctions; Linearity; Resistors; Standards development; Uncertainty; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.278631
  • Filename
    278631