DocumentCode
1071590
Title
Shielding stray magnetic fields of open high field MRI magnets
Author
Kakugawa, S. ; Hino, N. ; Komura, A. ; Kitamura, M. ; Takeshima, H. ; Yatsuo, T. ; Tazaki, H.
Author_Institution
Hitachi Res. Lab., Hitachi Ltd., Ibaraki, Japan
Volume
14
Issue
2
fYear
2004
fDate
6/1/2004 12:00:00 AM
Firstpage
1639
Lastpage
1642
Abstract
We have developed a new magnet for open high field (0.7 T) MRI systems. Since MRI systems are installed in hospital sites, it is important to reduce stray magnetic fields of the magnets. We studied passive, active and hybrid shielding methods and compared them. In order to study feasibility of active shielding, we considered higher order multipole stray fields. We present details of the calculation method and the results.
Keywords
biomedical MRI; linear programming; magnetic shielding; superconducting magnets; 0.7 T; MRI systems; active shielding; homogeneous magnetic fields; hospital sites; magnetic resonance imaging; medical diagnostics; multipole stray fields; open high field MRI magnets; open-type magnets; optimization; shielding methods; stray magnetic fields; Coils; Hospitals; Iron; Laplace equations; Magnetic flux; Magnetic resonance imaging; Magnetic shielding; Medical diagnosis; Saturation magnetization; Superconducting magnets; Linear programming; MRI; magnetic resonance imaging; multipole stray field; optimization;
fLanguage
English