DocumentCode
1078448
Title
Thin-film multielectrode arrays for a cochlear prosthesis
Author
Shamma-donoghue, Shihab A. ; May, Gerald A. ; Cotter, Neil E. ; White, Robert L. ; Simmons, F. Blair
Author_Institution
Stanford University, Stanford, CA
Volume
29
Issue
1
fYear
1982
fDate
1/1/1982 12:00:00 AM
Firstpage
136
Lastpage
144
Abstract
The design and fabrication of flexible thin-film microelectrode arrays for use in a cochlear prosthesis are described. The electrode array is designed to be inserted through the round window of the cochlea into the spiral scala tympani chamber of the cochlea. A lifetime of decades under stimulation is sought. The electrode array is comprised of photolithographically defined platinum-on-tantalum conductors sandwiched between polyimide layers. A liquid polyimide is used, which polymerizes in two stages. After the first stage of curing, the polyimide is susceptible to photolithographic etching, allowing patterned access holes to be cut into the top layer of the insulating sandwich. After the second cure, the polymer becomes inert biocompatible Kapton. The processing techniques and the electrode test results are presented.
Keywords
Conductors; Curing; Electrodes; Fabrication; Microelectrodes; Polyimides; Polymers; Prosthetics; Spirals; Transistors;
fLanguage
English
Journal_Title
Electron Devices, IEEE Transactions on
Publisher
ieee
ISSN
0018-9383
Type
jour
DOI
10.1109/T-ED.1982.20671
Filename
1482168
Link To Document