• DocumentCode
    1082663
  • Title

    Silicon ribbon cables for chronically implantable microelectrode arrays

  • Author

    Hetke, Jamille F. ; Lund, Jennifer L. ; Najafi, Khalil ; Wise, Kensall D. ; Anderson, David J.

  • Author_Institution
    Dept. of Electr. Eng. & Comput. Sci., Michigan Univ., Ann Arbor, MI, USA
  • Volume
    41
  • Issue
    4
  • fYear
    1994
  • fDate
    4/1/1994 12:00:00 AM
  • Firstpage
    314
  • Lastpage
    321
  • Abstract
    Describes the design, fabrication, and testing of miniature ultraflexible ribbon cables for use with micromachined silicon microprobes capable of chronic recording and/or stimulation in the central nervous system (CNS). These interconnects are of critical importance in reliably linking these microelectrodes to the external world through a percutaneous connector. The silicon cables allow the realization of multilead, multistrand shielded local interconnects that are extremely flexible and yet strong enough to withstand normal handling and surgical manipulation. Cables 5 μm thick, 1-5 cm long, and from 60 to 250 μm wide have been fabricated with up to eight leads. The series lead resistance is typically 4 kΩ/cm for polysilicon and 500 Ω/cm for tantalum with shunt capacitance values of 5-10 pF/cm and an interlead capacitance below 10 fF/cm. Soak tests in buffered saline performed under electrical and mechanical stress have been underway for over three years and show subpicoampere leakage levels. Silicon microprobes with built-in ribbon cables have remained functional for up to one gear in the guinea pig CNS, recording driven single-unit activity and maintaining impedance levels in the 1-7 M Ω range.
  • Keywords
    biological techniques and instruments; cables (electric); microelectrodes; neurophysiology; 1 to 5 cm; 1 to 7 Mohm; 3 y; 5 mum; 60 to 250 mum; Si; biophysical instrumentation; buffered saline; central nervous system; chronic recording; chronic stimulation; chronically implantable microelectrode arrays; electrical stress; guinea pig CNS; mechanical stress; miniature ultraflexible ribbon cables; multilead multistrand shielded local interconnects; percutaneous connector; polysilicon; shunt capacitance; silicon ribbon cables; Cables; Capacitance; Central nervous system; Connectors; Fabrication; Joining processes; Microelectrodes; Silicon; Surgery; System testing; Animals; Central Nervous System; Electric Impedance; Electrodes, Implanted; Equipment Design; Guinea Pigs; Materials Testing; Microelectrodes; Silicon; Stress, Mechanical; Time Factors;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.284959
  • Filename
    284959