• DocumentCode
    1096166
  • Title

    High-Accuracy Data Acquisition Architectures for Ultrasonic Imaging

  • Author

    Kalashnikov, Alexander N. ; Ivchenko, Vladimir G. ; Challis, Richard E. ; Hayes-Gill, Barrie R.

  • Author_Institution
    Sch. of Electr. & Electron. Eng., Nottingham Univ., Nottingham
  • Volume
    54
  • Issue
    8
  • fYear
    2007
  • fDate
    8/1/2007 12:00:00 AM
  • Firstpage
    1596
  • Lastpage
    1605
  • Abstract
    This paper proposes a novel architecture for a data acquisition system intended to support the next generation of ultrasonic imaging instruments operating at or above 100 MHz. Existing systems have relatively poor signal-to-noise ratios and are limited in terms of their maximum data sampling rate, both of which are improved by a combination of embedded averaging and embedded interleaved sampling. "On-the-fly" pipelined operation minimizes control overheads for signal averaging. A two-clock sampling timing system provides for effective sampling rates that are a factor of 20 or more above the basic sampling rate of the analog-to-digital converter (ADC). The system uses commercial field-programmable gate array devices operated at clock frequencies commensurable with the ADC clock. Implementation is via the Xilinx Xtreme digital signal processing development kit, available at low cost. Sample rates of up to 2160 MHz have been achieved in combination with up to 16384 coherent averages using the above-mentioned off-the-shelf hardware.
  • Keywords
    analogue-digital conversion; biomedical ultrasonics; data acquisition; digital signal processing chips; field programmable gate arrays; Xilinx Xtremc digital signal processing development kit; analog-to-digital converter; embedded averaging; embedded interleaved sampling; field-programmable gate array devices; frequency 2160 MHz; high-accuracy data acquisition; signal averaging; signal-to-noise ratios; two-clock sampling timing system; ultrasonic imaging; Analog-digital conversion; Clocks; Data acquisition; Field programmable analog arrays; Instruments; Sampling methods; Signal sampling; Signal to noise ratio; Timing; Ultrasonic imaging; Data Collection; Image Processing, Computer-Assisted; Signal Processing, Computer-Assisted; Software; Transducers; Ultrasonography;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2007.430
  • Filename
    4291509