    Penalized weighted least-squares image reconstruction for positron emission tomography

  • Author

    Fessler, Jeffrey A.

  • Author_Institution
    Med. Center, Michigan Univ., Ann Arbor, MI, USA
  • Volume
    13
  • Issue
    2
  • fYear
    1994
  • fDate
    6/1/1994 12:00:00 AM
  • Firstpage
    290
  • Lastpage
    300
  • Abstract
    Presents an image reconstruction method for positron-emission tomography (PET) based on a penalized, weighted least-squares (PWLS) objective. For PET measurements that are precorrected for accidental coincidences, the author argues statistically that a least-squares objective function is as appropriate, if not more so, than the popular Poisson likelihood objective. The author proposes a simple data-based method for determining the weights that accounts for attenuation and detector efficiency. A nonnegative successive over-relaxation (+SOR) algorithm converges rapidly to the global minimum of the PWLS objective. Quantitative simulation results demonstrate that the bias/variance tradeoff of the PWLS+SOR method is comparable to the maximum-likelihood expectation-maximization (ML-EM) method (but with fewer iterations), and is improved relative to the conventional filtered backprojection (FBP) method. Qualitative results suggest that the streak artifacts common to the FBP method are nearly eliminated by the PWLS+SOR method, and indicate that the proposed method for weighting the measurements is a significant factor in the improvement over FBP
  • Keywords
    computerised tomography; image reconstruction; medical image processing; radioisotope scanning and imaging; Poisson likelihood objective; accidental coincidences; attenuation; bias/variance tradeoff; detector efficiency; filtered backprojection method; image reconstruction method; maximum-likelihood expectation-maximization method; medical diagnostic imaging; nonnegative successive over-relaxation algorithm; nuclear medicine; penalized weighted least-squares image reconstruction; positron emission tomography; simple data-based method; Attenuation; Detectors; Humans; Image reconstruction; Physiology; Positron emission tomography; Reconstruction algorithms; Statistics; US Department of Energy; Weight measurement;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/42.293921
  • Filename