• DocumentCode
    1142867
  • Title

    Performance of X-Ray Imaging Systems Applied to Intravenous Angiography

  • Author

    Riederer, S.J.

  • Author_Institution
    General Electric Co., Medical Systems, P. O. Box 414, NB-924, Milwaukee, WI 53201
  • Volume
    1
  • Issue
    1
  • fYear
    1982
  • fDate
    7/1/1982 12:00:00 AM
  • Firstpage
    48
  • Lastpage
    62
  • Abstract
    A review is made of the performance of X-ray systems applied to the imaging of dilute amounts of iodine using subtraction methods. Relevant performance criteria are defined and desired performance levels are derived. For a general purpose system these are iodine sensitivity of 1 mg/cm2, cancellation of 10 g/cm2 soft tissue and 2 g/cm2 bone to be equivalent to iodine sensitivity, capability of imaging through 25 cm soft tissue, spatial resolution exceeding 1 lp/mm and acceptable imaging time depending on the application. Conceptual aspects of several subtraction techniques are discussed including temporal, generalized temporal, dual-energy K-edge, dual-energy non-K-edge, three-energy K-edge, and hybrid subtraction. Experimental performance of several representative area and linear detector systems is tabulated according to subtraction algorithm. A table of references is provided.
  • Keywords
    Angiography; Biological tissues; Biomedical imaging; Bones; Image resolution; Object detection; Optical imaging; X-ray detection; X-ray detectors; X-ray imaging;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.1982.4307548
  • Filename
    4307548