• DocumentCode
    1143033
  • Title

    Reconstruction from Sparsely Sampled Data by ART with Interpolated Rays

  • Author

    Kouris, K. ; Tuy, H. ; Lent, A. ; Herman, G.T. ; Lewitt, R.M.

  • Volume
    1
  • Issue
    3
  • fYear
    1982
  • Firstpage
    161
  • Lastpage
    167
  • Abstract
    After a brief discussion of the algebraic reconstruction techniques (ART), we introduce the attenuation problem in positron emission tomography (PET). We anticipate that a generalization of ART, the so-called cyclic subgradient projection (CSP) method, may be useful for solving this problem. This, however, has not been successfully realized, due to the fact that data collected by our proposed stationary PET detector ring are too sparsely sampled. That this is, in fact, a major problem is demonstrated by showing that ordinary ART produces reconstructions with unacceptably strong artifacts even on perfect (no attenuation) data collected according to the PET geometry. We demonstrate that the source of this artifact is the sparse sampling, and we propose the use of interpolated rays to overcome the problem. This approach is successful, as is illustrated by showing reconstructions from sparsely sampled data by ART with interpolated rays.
  • Keywords
    Attenuation; Biomedical imaging; Detectors; Equations; Image reconstruction; Pixel; Positron emission tomography; Radiology; Subspace constraints; Vectors;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.1982.4307567
  • Filename
    4307567