• DocumentCode
    1143448
  • Title

    A Regularizing Method for Quantitative SPECT Reconstruction

  • Author

    Soussaline, F. ; Le Coq, G.

  • Volume
    2
  • Issue
    1
  • fYear
    1983
  • fDate
    3/1/1983 12:00:00 AM
  • Firstpage
    24
  • Lastpage
    30
  • Abstract
    The problem of successive transverse plane reconstruction in single photon emission computerized tomography (SPECT) is modeled in its more general form, which implies the definition of emission tomographic operators (ETO´s) for which an analytical solution can be derived. The properties of the attenuated tomographic operator (ATO) are described and discussed, including the attenuation which is distributed on the reconstruction domain. For this particular operator, a regularizing method (RIM) is proposed, for which it is demonstrated and tested with simulation studies that a filtered, accurate solution can be extracted for the tomographic images as obtained using a single photon emission tomograph based on a rotating gamma camera in clinical use.
  • Keywords
    Attenuation; Collimators; Computed tomography; Data mining; Detectors; Extraterrestrial measurements; Image reconstruction; Low pass filters; Single photon emission computed tomography; X-ray imaging;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.1983.4307609
  • Filename
    4307609