• DocumentCode
    1144513
  • Title

    High-Speed Spiral-Scan Echo Planar NMR Imaging-I

  • Author

    Ahn, C.B. ; Kim, J.H. ; Cho, Z.H.

  • Volume
    5
  • Issue
    1
  • fYear
    1986
  • fDate
    3/1/1986 12:00:00 AM
  • Firstpage
    2
  • Lastpage
    7
  • Abstract
    An improved echo planar high-speed imaging technique using spiral scan is presented and experimental advantages are discussed. This proposed spiral-scan echo planar imaging (SEPI) technique employs two linearly increasing sinusoidal gradient fields, which results in a spiral trajectory in the spatial frequency domain (k-domain) that covers the entire frequency domain uniformly. The advantages of the method are: 1) circularly symmetric T2 weighting, resulting in a circularly symmetric point spread function in the image domain; 2) elimination of discontinuities in gradient waveforms which in turn will reduce initial transient as well as steady-state distortions; and 3) effective rapid spiral-scan from dc to high frequency in a continuous fashion, which ensures multiple pulsing with interlacing for further resolution improvement without T2 decay image degradation. Some preliminary experimental results will be presented and further possible improvements suggested.
  • Keywords
    Frequency domain analysis; High-resolution imaging; Image reconstruction; Image resolution; Magnetic resonance imaging; Nuclear magnetic resonance; Optical imaging; Sampling methods; Spirals; Student members;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.1986.4307732
  • Filename
    4307732