DocumentCode
1144562
Title
Shading 3D-Images from CT Using Gray-Level Gradients
Author
Hohne, Karl Heinz ; Bernstein, Ralph
Volume
5
Issue
1
fYear
1986
fDate
3/1/1986 12:00:00 AM
Firstpage
45
Lastpage
47
Abstract
For the 3D-reconstruction of organ surfaces from tomograms, a shading method based on the partial volume effect is presented. In contrast to methods based on the depth and/or the angle of the voxel surface, here the gray-level gradient along the surface is used for shading. It is shown, that at least for bone and soft tissue surfaces, the results are superior to conventional shading. This is due to the high dynamic range of the gray levels within a small spatial neighborhood.
Keywords
Biological tissues; Bones; Computed tomography; Dynamic range; Head; Interpolation; Layout; Quantization; Surface structures; Surgery;
fLanguage
English
Journal_Title
Medical Imaging, IEEE Transactions on
Publisher
ieee
ISSN
0278-0062
Type
jour
DOI
10.1109/TMI.1986.4307738
Filename
4307738
Link To Document