• DocumentCode
    1189675
  • Title

    An International Comparison of Inductive Voltage Divider Calibrations at 400 and 1000 Hz

  • Author

    Sze, W.C. ; Dunn, A.F. ; Zapf, T.L.

  • Issue
    3
  • fYear
    1965
  • Firstpage
    124
  • Lastpage
    131
  • Abstract
    The establishment and maintenance of the highest possible order of accuracy for the measurements of voltage ratio and phase angle are responsibilities of the national laboratories. This paper reports the results of an international comparison of a seven-decade inductive voltage divider by the United States National Bureau of Standards and the National Research Council of Canada. The calibration techniques used in each laboratory are discussed in detail, and the results of calibration at 400 and 1000 Hz are compared. The accuracies of measurements are within 0.1 ppm of input for voltage ratio and 1 (setting of dials)1/2 microradians for phase angle.
  • Keywords
    Bridge circuits; Calibration; Circuit testing; Councils; Frequency; Laboratories; Magnetic cores; NIST; Phase measurement; Voltage measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1965.4313459
  • Filename
    4313459