DocumentCode
1190071
Title
Evaluation of a Thallium Atomic Beam Frequency Standard at the National Bureau of Standards
Author
Beehler, R.E. ; Glaze, D.J.
Volume
15
Issue
42371
fYear
1966
Firstpage
55
Lastpage
58
Abstract
The original NBS cesium standard (NBS I) has been converted to a thallium standard and was operated for one and one-half years with a typical precision of 2 Ã 10-12 and an accuracy of 1 Ã 10-11. Experiments are described which were performed to establish these precision and accuracy estimates. These results, which are comparable to those obtained with longer cesium standards, are considered sufficiently encouraging to justify the conversion of a longer cesium standard to thallium for a more thorough evaluation.
Keywords
Atomic beams; Frequency measurement; Ionization; Leak detection; Measurement standards; NIST; Temperature; Tungsten; Uncertainty; Wire;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1966.4313502