• DocumentCode
    1197394
  • Title

    Extensions of Time Domain Metrology above 10 GHz to Materials Measurements

  • Author

    Cronson, Harry M. ; Nicolson, Murray A. ; Mitchell, Peter G.

  • Volume
    23
  • Issue
    4
  • fYear
    1974
  • Firstpage
    463
  • Lastpage
    468
  • Abstract
    A new time domain measurement technique to obtain the relative complex permeability , ¿* and permittivity ¿* of materials at frequencies above 10 GHz is described. The main feature is the use of a 400-ps-wide RF pulse-excitation waveform that furnishes the necessary spectral amplitude in the 9-to 16-GHz frequency range. The delay line configuration and instrumentation are similar to a previously reported lower frequency system where a sample of solid material less than 5 mm thick is cut to fit into a coaxial sample holder. Experimental values of ¿* and ¿* are given for Teflon and a ferrite material for measurements made in two segments spanning 0.4-10 GHz and 9-16 GHz. These show good agreement with data obtained from direct frequency domain measurements. A discussion is included on the amplitude spectra of a step-recovery diode and tunnel diode pulse generator and their application as time domain sources. Considerations for the optimum choice of sample thickness are also given.
  • Keywords
    Delay lines; Diodes; Instruments; Measurement techniques; Metrology; Permeability; Permittivity measurement; Radio frequency; Solids; Time measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1974.4314335
  • Filename
    4314335