DocumentCode
1203000
Title
60-Hz Interference in Electrocardiography
Author
Huhta, James C. ; Webster, John G.
Author_Institution
Department of Electrical Engineering, University of Wisconsin, Madison, Wis. 53706.
Issue
2
fYear
1973
fDate
3/1/1973 12:00:00 AM
Firstpage
91
Lastpage
101
Abstract
One of the major problems encountered in recording ECG´s is the appearance of unwanted 60-Hz interference in the output. This paper examines the many possible sources of interference, and for each provides a description, an identifying test, and a remedy. Then we look at possible equipment imperfections and guidelines for amplifier design to aid in the evaluation of new developments in the field of electrocardiography. As an illustrative example, we apply the procedure for isolating the source of interference to an exposition display in which each visitor could rapidly view his own ECG.
Keywords
Electric fields; Electrocardiography; Interference elimination; Magnetic recording; Magnetic separation; Magnetic shielding; Radio frequency; Switches; Testing; Wires; Biomedical Engineering; Data Display; Electric Conductivity; Electrocardiography; Electromagnetics; Electronics, Medical; Mathematics; Tape Recording;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.1973.324169
Filename
4120603
Link To Document