• DocumentCode
    1203368
  • Title

    An Automatic System for AC/DC Calibration

  • Author

    Lentner, K.J. ; Flach, Donald R.

  • Volume
    32
  • Issue
    1
  • fYear
    1983
  • fDate
    3/1/1983 12:00:00 AM
  • Firstpage
    51
  • Lastpage
    56
  • Abstract
    An automatic ac/dc difference calibration system using direct measurement of thermoelement EMFs is described. The system operates over a frequency range from 20 Hz to 100 kHz, covering the voltage range from 0.5 V to 1 kV. For all voltages, the total uncertainty (including the uncertainty of the specific reference thermal converters used) is 50 ppm at frequencies from 20 Hz to 20 kHz, inclusive, and 100 ppm at higher frequencies up to 100 kHz. In addition to ac/dc difference testing, the system can be used to measure some important characteristics of thermoelements, as well as to calibrate ac voltage calibrators and precision voltmeters. Results of intercomparisons between the new system and the manual NBS calibration system, using single-range, coaxial-type, thermal voltage converters as transfer standards, are reported. The results indicate that the ac/dc differences measured are accurate to well within the combined total uncertainty limits of the two systems.
  • Keywords
    Automatic testing; Calibration; Frequency conversion; NIST; Performance evaluation; System testing; Tellurium; Uncertainty; Voltage; Voltmeters;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1983.4315009
  • Filename
    4315009