DocumentCode
1203368
Title
An Automatic System for AC/DC Calibration
Author
Lentner, K.J. ; Flach, Donald R.
Volume
32
Issue
1
fYear
1983
fDate
3/1/1983 12:00:00 AM
Firstpage
51
Lastpage
56
Abstract
An automatic ac/dc difference calibration system using direct measurement of thermoelement EMFs is described. The system operates over a frequency range from 20 Hz to 100 kHz, covering the voltage range from 0.5 V to 1 kV. For all voltages, the total uncertainty (including the uncertainty of the specific reference thermal converters used) is 50 ppm at frequencies from 20 Hz to 20 kHz, inclusive, and 100 ppm at higher frequencies up to 100 kHz. In addition to ac/dc difference testing, the system can be used to measure some important characteristics of thermoelements, as well as to calibrate ac voltage calibrators and precision voltmeters. Results of intercomparisons between the new system and the manual NBS calibration system, using single-range, coaxial-type, thermal voltage converters as transfer standards, are reported. The results indicate that the ac/dc differences measured are accurate to well within the combined total uncertainty limits of the two systems.
Keywords
Automatic testing; Calibration; Frequency conversion; NIST; Performance evaluation; System testing; Tellurium; Uncertainty; Voltage; Voltmeters;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.1983.4315009
Filename
4315009
Link To Document