Title
Induction Thermocoagulation - A Seed Power Study
Author
Merry, G.Allen ; Hale, Richard ; Zervas, Nicholas T.
Issue
4
fYear
1973
fDate
7/1/1973 12:00:00 AM
Firstpage
302
Lastpage
303
Abstract
Thermal-power output of destructive therapeutic implants composed of nickel-palladium Curie alloys or 430 stainless steel (a few watts in typical use) is presented. Eddy-current models of implant heating are discussed. A field strength of about 5000 A/m-1 is shown to be appropriate for seeds of about 2-mm OD.
Keywords
Eddy currents; Equations; Frequency; Heating; Implants; Neurosurgery; Permeability; Production; Skin; Temperature measurement; Calorimetry; Heat; Methods; Neurosurgery; Thermodynamics;