• DocumentCode
    1203726
  • Title

    Performance of the Four NRC Long-Beam Primary Cesium Clocks Cs V, Cs VIA, Cs VIB, and Cs VIC

  • Author

    Mungall, Allan G. ; Costain, Cecil C.

  • Volume
    32
  • Issue
    1
  • fYear
    1983
  • fDate
    3/1/1983 12:00:00 AM
  • Firstpage
    224
  • Lastpage
    227
  • Abstract
    Recent performance of the four laboratory primary cesium clocks at the National Research Council (NRC) is outlined and the results of both internal and external time scale comparisons described. Difficulties which have been experienced with these clocks, including failures in both the microwave resonance excitation systems and the vacuum systems are discussed. The results of seven years of primary clock operation for Cs V and over two years for the three Cs VI clocks show that all the clocks have operated well within their estimated accuracy and stability limits of ±1 × 10-13 for Cs V and ±1.5 × 10-13 for Cs VIA, Cs VIB, and Cs VIC.
  • Keywords
    Clocks; Councils; Frequency; Laboratories; NIST; Observatories; Resonance; Satellites; Stability; Vacuum systems;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1983.4315046
  • Filename
    4315046