    60-Hz Ventricular Fibrillation and Pump Failure Thresholds Versus Electrode Area

  • Author

    Roy, O.Z. ; Scott, John R. ; Park, Gordon C.

  • Author_Institution
    Radio and Electrical Engineering Division, National Research Council of Canada, Ottawa, Ont., Canada.
  • Issue
    1
  • fYear
    1976
  • Firstpage
    45
  • Lastpage
    48
  • Abstract
    Fibrillation thresholds and safe leakage current levels are being reexamined and new values of safe current other than the 10 ¿A are being considered. Values of fibrillation and pump failure thresholds are presented and related to surface area. The surface areas covered range from those found in dye injection and other exploratory catheters (0.224 mm2) to the largest endocardial pacing catheter (90 mm2). The results indicate that to choose or set a safe current level without regard to surface area or based on one surface area is unrealistic.
  • Keywords
    Angiography; Catheters; Councils; Electrodes; Fibrillation; Heart; Helium; Leakage current; Mercury (metals); Pacemakers; Animals; Dogs; Electrodes; Heart Catheterization; Heart Failure; Pacemaker, Artificial; Surface Properties; Ventricular Fibrillation;