• DocumentCode
    1208661
  • Title

    Reevaluation of the Optically Pumped Cesium Frequency Standard NRLM-4 With an H-Bend Ring Cavity

  • Author

    Hagimoto, Ken ; Koga, Yasuki ; Ikegami, Takeshi

  • Author_Institution
    Nat. Inst. of Adv. Ind. Sci. & Technol., Metrol. Inst. of Japan, Tsukuba
  • Volume
    57
  • Issue
    10
  • fYear
    2008
  • Firstpage
    2212
  • Lastpage
    2217
  • Abstract
    The largest source of uncertainty of the optically pumped cesium (Cs) frequency standard (NRLM-4), i.e., the distributed cavity phase shift, was reduced by replacing the microwave cavity with a new ring-type one. An H-bend-type ring cavity is adopted for the first time. Due to the novel cavity and the improvement of the Cs beam alignment, the distributed cavity phase shift was significantly reduced from 2 x 10-14 to 1.4 x 10-15. The frequency stability was measured to be 8 x 10-13 tau-1/2 . The total uncertainty was reevaluated to be 6.7 x 10-15, which is four times better than the previous value.
  • Keywords
    caesium; laser cavity resonators; optical pumping; H-bend-type ring cavity; frequency stability; microwave cavity; optically pumped cesium frequency standard; Beam type; H-bend-type ring cavity; International Atomic Time (TAI); optically pumped cesium (Cs) frequency standard; uncertainty;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2008.922086
  • Filename
    4509486