DocumentCode
1209215
Title
A New Data-Reduction Algorithm for Real-Time ECG Analysis
Author
Abenstein, John P. ; Tompkins, Willis J.
Author_Institution
University of Wisconsin, Madison, WI 53706. He is now with the Department of Anesthesia, Michael Reese Hospital and Medical Center
Issue
1
fYear
1982
Firstpage
43
Lastpage
48
Abstract
Typically the ECG is sampled at a rate of 200 samples/s or more, producing a large amount of data that are difficult to store, analyze, and transmit. Data-reduction algorithms that operate in real time reduce he amount of data without losing the clinical information content. They must also leave sufficient computation time available for ECG analysis. We describe here a new algorithm called CORTES that is suited for such real-time applications. This algorithm combines the best features of two other techniques called TP and AZTEC. We present the results of a study to find optimal experimental values for the controlling variables in CORTES. We compare the computations of root-mean-square reconstruction errors for a diversity of encoded ECG signals.
Keywords
Algorithm design and analysis; Bandwidth; Clinical diagnosis; Electric variables control; Electrocardiography; Frequency; Memory management; Optimal control; Signal analysis; Signal design; Computers; Electrocardiography; Humans; Time Factors;
fLanguage
English