• DocumentCode
    1211956
  • Title

    Transventricular Impedance During Fibrillation

  • Author

    Savino, Guillermo V. ; Del Valle Ruiz, Estela ; Valentinuzzi, Max E.

  • Author_Institution
    Departamento Bioingenieria, INSIBIO, Universidad Nacional de Tucuman
  • Issue
    6
  • fYear
    1983
  • fDate
    6/1/1983 12:00:00 AM
  • Firstpage
    364
  • Lastpage
    367
  • Abstract
    Transventricular impedance during fibrillation was measured with two methods: 1) by injecting a low constant sinusoidal current (obtaining a Z value), and 2) by using the time constant of an exponential defibrillatory discharge (giving an R value). One group of eight dols (Group 300) was normothermic and another group of 14 dogs (Group 400) was hypothermic. The left anterior descending coronary artery was ligated in both groups. For Group 300 (values in ohms), Z = 29.5 (SD 5.1), R = 30.9 (SD 4.7). The paired t-test gave a statistically significant difference (p 0.1 percent). For Group 400, Z = 34.5 (SD 7.3); R = 33.8 (SD 5.8), also with a significant difference, but at a lower level (p 5 percent). Nonetheless, from the practical point of view of calculating the voltage to be applied after establishing an adequate dose current, they (the differences) do not seem to have any important quantitative weight, supporting method 1 (for its easiness) to estimate transventricular impedance in this particular application.
  • Keywords
    Delay; Diseases; Dispersion; Electroencephalography; Electromyography; Fast Fourier transforms; Impedance; Multiple sclerosis; Muscles; Nervous system; Animals; Dogs; Electric Conductivity; Electric Countershock; Female; Heart; Male;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1983.325140
  • Filename
    4121650