• DocumentCode
    1212458
  • Title

    Calibration of electronic capacitance and dissipation factor bridges

  • Author

    Ramm, Günther ; Moser, Harald

  • Author_Institution
    Phys. Tech. Bundesanstalt, Braunschweig, Germany
  • Volume
    52
  • Issue
    2
  • fYear
    2003
  • fDate
    4/1/2003 12:00:00 AM
  • Firstpage
    396
  • Lastpage
    399
  • Abstract
    A new calibration procedure for electronic capacitance and dissipation factor bridges is presented. These bridges measure the ratio of their input currents. For the calibration, two sinusoidal alternating currents are synthesized with the aid of a digital, programmable two-channel AC voltage source and two AC resistors. The complex ratio of these synthesized input currents is calculated from an AC voltage ratio determined by a sampling method, together with the impedance ratio of the AC resistors. At currents of a few mA, their complex ratio in the frequency range up to 100 Hz can be measured with uncertainties (k=2) of approximately 1×10-6 and 1 μrad, respectively.
  • Keywords
    bridge circuits; bridge instruments; calibration; capacitance measurement; electric impedance; measurement uncertainty; programmable circuits; signal sampling; voltage measurement; 100 Hz; AC bridge; AC resistor impedance ratio; calibration; complex voltage ratio; digital programmable two-channel AC voltage source; electronic capacitance/dissipation factor bridges; frequency range; impedance ratio; input current ratio measurement; measurement uncertainties; sampling voltmeter; sinusoidal alternating current synthesis; Bridge circuits; Calibration; Capacitance; Current measurement; Frequency measurement; Impedance; Measurement uncertainty; Resistors; Sampling methods; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2003.810719
  • Filename
    1202056