DocumentCode
1216378
Title
Criteria for Optimal Averaging of Cardiac Signals
Author
Craelius, William ; Restivo, Mark ; Assadi, Mahshid A. ; El-Sherif, Nabil
Author_Institution
Cardiology Department, Veterans Administration Medical Center and S. U. N. Y. Health Science Center
Issue
10
fYear
1986
Firstpage
957
Lastpage
966
Abstract
The averaging process is modeled as a linear system whose low-pass filter characteristics are determined by the degree in temporal misalignment of signals. Assuming the errors in temporal alignment of successive cardiac cycles are random, then the model transfer function is equivalent to the probability density function. The response of the model to a step input is equivalent to the probability distribution function, which can be readily quantified. To validate the model, a high resolution ECG amplifier and QRS recognition system was constructed that synchronizes a step input with a point on the QRS. Design criteria for optimal amplification, filtering, and triggering of the ECG are determined. Test of the model reveals a close correspondence between observed and predicted step responses. From the average step response, the recording fidelity of any average can be determined-rapidly while the alignment is adjusted for optimal precision. Using ECG signals from patients, our model system demonstrates that alignment errors can both add and subtract signal components. Methods for estimating the extent of signal distortion induced by averaging as well as criteria for minimizing it are presented.
Keywords
Electrocardiography; Filtering; Linear systems; Low pass filters; Nonlinear filters; Probability density function; Probability distribution; Signal processing; Testing; Transfer functions; Biomedical Engineering; Biometry; Electrocardiography; Humans; Signal Processing, Computer-Assisted;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.1986.325667
Filename
4122183
Link To Document