• DocumentCode
    1218307
  • Title

    A Reliable Myoelectric Signal Detector Based on the Propagation Characteristics of Motor Unit Action Potentials

  • Author

    Masuda, Tadashi

  • Author_Institution
    Industrial Products Research Institute, Tsukuba Science City
  • Issue
    9
  • fYear
    1986
  • Firstpage
    876
  • Lastpage
    878
  • Abstract
    A method was proposed for detecting and rejecting motion artifacts superimposed on myoelectric (ME) signals which are used in the estimation of muscular activity, in the control of powered prostheses, and in other applications. The method is based on the propagation characteristics of motor unit action potentials derived with multiple surface electrodes placed along the muscle fibers. The contamination of artifacts was detected by the decrease of the normalized correlation coefficient calculated at the time shift corresponding to the potential propagation. The product of two correlated signals was found to be less affected by the artifacts and was a better estimate of muscular activity than the root mean square of the ME signal which is conventionally used in the applications of ME signals.
  • Keywords
    Detectors; Electrodes; Motion control; Motion detection; Motion estimation; Muscles; Prosthetics; Root mean square; Signal detection; Surface contamination; Adult; Electromyography; Humans; Male; Motor Neurons; Muscles; Neural Conduction;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1986.325782
  • Filename
    4122413