• DocumentCode
    1221739
  • Title

    Error Analysis for Vector Scope Measurement of Pulmonary Mechanics

  • Author

    Koen, Peter A. ; Moskowitz, Gordon D. ; Shaffer, Thomas H.

  • Author_Institution
    Department of Physiology, University of Pennsylvania
  • Issue
    2
  • fYear
    1978
  • fDate
    3/1/1978 12:00:00 AM
  • Firstpage
    204
  • Lastpage
    205
  • Abstract
    The errors that are incurred by misreading the angle on the oscilloscope face in the determination of pulmonary compliance, resistance, and functional residual capacity with a vector scope are presented. Results indicate that the calculated value will be within 10% of the correct value, provided that the oscilloscope angle is between 10.5 and 78.5 degrees, and the reading error is less than 1 degree.
  • Keywords
    Biomedical measurements; Calibration; Electrical resistance measurement; Equations; Error analysis; Immune system; Lungs; Measurement errors; Mechanical variables measurement; Oscilloscopes; Humans; Plethysmography; Respiration;