DocumentCode
1221739
Title
Error Analysis for Vector Scope Measurement of Pulmonary Mechanics
Author
Koen, Peter A. ; Moskowitz, Gordon D. ; Shaffer, Thomas H.
Author_Institution
Department of Physiology, University of Pennsylvania
Issue
2
fYear
1978
fDate
3/1/1978 12:00:00 AM
Firstpage
204
Lastpage
205
Abstract
The errors that are incurred by misreading the angle on the oscilloscope face in the determination of pulmonary compliance, resistance, and functional residual capacity with a vector scope are presented. Results indicate that the calculated value will be within 10% of the correct value, provided that the oscilloscope angle is between 10.5 and 78.5 degrees, and the reading error is less than 1 degree.
Keywords
Biomedical measurements; Calibration; Electrical resistance measurement; Equations; Error analysis; Immune system; Lungs; Measurement errors; Mechanical variables measurement; Oscilloscopes; Humans; Plethysmography; Respiration;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.1978.326250
Filename
4122807
Link To Document