• DocumentCode
    1234139
  • Title

    Pulsed phase-locked loop calibration over frequency

  • Author

    Froggatt, Mark E. ; Allison, Sidney G. ; Moore, Jason P.

  • Author_Institution
    NASA Langley Res. Center, Hampton, VA, USA
  • Volume
    45
  • Issue
    1
  • fYear
    1996
  • fDate
    2/1/1996 12:00:00 AM
  • Firstpage
    123
  • Lastpage
    126
  • Abstract
    This paper presents a method of correcting for unwanted phase shifts introduced by interchanging ultrasonic transducers and other measurement system components when using the pulsed phase-locked loop (PPLL) ultrasonic system. Theory is derived mathematically separating phase errors into their constituents. The relationship to a previously derived method of determining the number of periods into the tone-burst, at which the system samples, is established, and a method of adjusting this number to be an integer is presented. Implementation of the technology on a computer is demonstrated through experimental results on transducers of differing frequencies and diameters. Sensitivity to path-length is tested. This work has resulted in a straightforward means of calibrating transducers in conjunction with PPLL measurements using a single, fixed reference block
  • Keywords
    calibration; computerised instrumentation; error correction; measurement errors; phase locked loops; ultrasonic measurement; ultrasonic transducers; calibration; diameters; fixed reference block; path-length; phase errors; pulsed phase-locked loop; tone-burst; ultrasonic transducers; unwanted phase shifts; Acoustic measurements; Bonding; Calibration; Frequency locked loops; Phase detection; Phase locked loops; Phase measurement; Pulse measurements; Transducers; Ultrasonic variables measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.481323
  • Filename
    481323