• DocumentCode
    1234479
  • Title

    A 1000 A/20 kV/25 kHz-500 kHz Volt-Ampere-Wattmeter for loads with power factors from 0.001 to 1.00

  • Author

    Clarke, Kenneth K. ; Hess, Donald T.

  • Author_Institution
    Clarke-Hess Commun. Res. Corp., New York, NY, USA
  • Volume
    45
  • Issue
    1
  • fYear
    1996
  • fDate
    2/1/1996 12:00:00 AM
  • Firstpage
    142
  • Lastpage
    145
  • Abstract
    Many industrial processes and scientific experiments utilize large amounts of ac power at frequencies from 3 kHz to 500 kHz. The phase angle and the impedances of these loads often vary over a wide range. This paper describes an instrument to provide an accurate measurement of currents (1 A to 1000 A), voltages (100 V to 20 kV), and powers (100 W to 20 MW) over the frequency range from 25 kHz to 500 kHz. It deals with loads having power factors down to nearly zero and with load impedances from 10 Ω to 20 kΩ. The paper contains practical details of both the design and the calibration of the front-end voltage and current transducers. The characteristics of a nearly ideal broadband current transducer are presented. Overall instrument calibration, verification and traceability problems are considered in detail
  • Keywords
    calibration; electric current measurement; power measurement; transducers; voltage measurement; wattmeters; 1 to 1000 A; 10 ohm to 20 kohm; 100 V to 20 kV; 25 to 500 kHz; ac power; broadband current transducer; calibration; capacitive attenuator; design; front-end voltage transducers; impedances; industrial processes; phase angle; scientific experiments; traceability; verification; Calibration; Frequency; Impedance; Instruments; Plasma applications; Plasma measurements; Reactive power; Transducers; Voltage; Wattmeters;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.481326
  • Filename