• DocumentCode
    1244546
  • Title

    Determination of mechanical and electronic shifts for pinhole SPECT using a single point source

  • Author

    Metzler, S.D. ; Greer, K.L. ; Jaszczak, R.J.

  • Author_Institution
    Dept. of Radiol., Pennsylvania Univ., Philadelphia, PA, USA
  • Volume
    24
  • Issue
    3
  • fYear
    2005
  • fDate
    3/1/2005 12:00:00 AM
  • Firstpage
    361
  • Lastpage
    370
  • Abstract
    The effects of uncompensated electronic and mechanical shifts may compromise the resolution of pinhole single photon emission computed tomography. The resolution degradation due to uncompensated shifts is estimated through simulated data. A method for determining the transverse mechanical and axial electronic shifts is described and evaluated. This method assumes that the tilt of the detector and the radius of rotation (ROR) are previously determined using another method. When this assumption is made, it is possible to determine the rest of the calibration parameters using a single point source. A method that determines the electronic and mechanical shifts as well as the tilt has been previously described; this method requires three point sources. It may be reasonable in most circumstances to calibrate tilt much less frequently than the mechanical shifts since the tilt is a property of the scanner whereas the mechanical shift may change every time the collimator is replaced. An alternative method for determining the ROR may also be used. Lastly, we take the view that the transverse electronic shift and the focal length change slowly and find these parameters independently.
  • Keywords
    calibration; image resolution; medical image processing; single photon emission computed tomography; axial electronic shift; calibration parameters; collimator; focal length change; pinhole SPECT; pinhole single photon emission computed tomography; resolution degradation; single point source; transverse electronic shift; transverse mechanical shift; uncompensated shifts; Biomedical engineering; Biomedical imaging; Calibration; Collimators; Computational modeling; Degradation; Detectors; Image reconstruction; Radiology; Single photon emission computed tomography; Calibration; SPECT; image reconstruction; pinhole; single photon emission computed tomography; Algorithms; Artifacts; Calibration; Electromagnetic Fields; Equipment Failure Analysis; Image Enhancement; Image Interpretation, Computer-Assisted; Imaging, Three-Dimensional; Mechanics; Phantoms, Imaging; Reproducibility of Results; Sensitivity and Specificity; Tomography, Emission-Computed, Single-Photon;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.2004.842456
  • Filename
    1397823