• DocumentCode
    1252847
  • Title

    Experimental validation of a backpropagation algorithm for three-dimensional breast tumor localization

  • Author

    Liu, Hanli ; Matson, Charles L. ; Lau, Kelly ; Mapakshi, Rajesh R.

  • Author_Institution
    Biomed. Eng. Program, Texas Univ., Arlington, TX, USA
  • Volume
    5
  • Issue
    4
  • fYear
    1999
  • Firstpage
    1049
  • Lastpage
    1057
  • Abstract
    To validate a new backpropagation algorithm for three-dimensional (3-D) tumor localization in breast tissue with a two-dimensional measurement, we performed experiments on breast phantoms using a heterodyne, frequency-domain photon migration system. The near-infrared (680 or 780 nm) laser diode used to provide illumination was intensity modulated at both megahertz (20-60 MHz) and kilohertz (10 kHz) frequencies. The breast phantoms were made of plastic resin with a variety of simulated tumors imbedded inside and then scanned with the system in a planar geometry. With megahertz modulation signals, both of the amplitude and phase of the transmitted light were used in the data reduction process, whereas with kilohertz modulation signals only the amplitudes were used for the reconstruction. In both cases, the backpropagation reconstruction algorithm was used to accomplish 3-D localization of the imbedded tumors. In all cases, the reconstructed locations for the hidden objects are in good agreement with the true values. Our current work demonstrates the possibility and potential of developing low-cost optical tomographic instruments
  • Keywords
    backpropagation; image reconstruction; laser applications in medicine; mammography; medical image processing; optical tomography; tumours; 680 nm; 780 nm; backpropagation reconstruction algorithm; breast phantoms; breast tissue; data reduction process; heterodyne frequency-domain photon migration system; hidden objects; imbedded tumor; kilohertz modulation signals; low-cost optical tomographic instruments; megahertz modulation signals; near-infrared laser diode; planar geometry; plastic resin; reconstructed locations; simulated tumors; three-dimensional breast tumor localization; two-dimensional measurement; Amplitude modulation; Backpropagation algorithms; Breast neoplasms; Breast tissue; Frequency measurement; Imaging phantoms; Optical modulation; Performance evaluation; Phase modulation; Signal processing;
  • fLanguage
    English
  • Journal_Title
    Selected Topics in Quantum Electronics, IEEE Journal of
  • Publisher
    ieee
  • ISSN
    1077-260X
  • Type

    jour

  • DOI
    10.1109/2944.796329
  • Filename
    796329