• DocumentCode
    129454
  • Title

    Performance optimization of ultrasonic needle actuating device for insertion operation into tissue mimics

  • Author

    Xiaochun Liao ; Sadiq, Muhammad ; Yang Kuang ; Corner, George ; Cochran, Sandy ; Zhihong Huang

  • Author_Institution
    Sch. of Eng., Phys. & Math., Univ. of Dundee, Dundee, UK
  • fYear
    2014
  • fDate
    3-6 Sept. 2014
  • Firstpage
    823
  • Lastpage
    826
  • Abstract
    Needle insertion with ultrasound actuation has been proven as an effective way to reduce the penetration force and needle deflection within soft tissue, facilitating operational manoeuverability and targeting accuracy in cancer biopsy and regional anaesthesia. In such percutaneous procedures, loading conditions from adjacent soft tissue alter the needle transducer´s impedance and cause deterioration of its vibration performance. In this paper, we report the needle transducer´s structure and working principle; we investigate the loading effects from tissue mimics on the transducer´s resonant impedance and vibration performance; and we show how to enhance the needle transducer´s vibration performance through optimization of geometrical dimensions and the electrical driving method. Structural optimization was carried out with different transversal dimensions of the needle. Electrical driving optimization was done by tracking the working resonant frequency and stabilizing the driving signal during operation. Experimental trials confirmed a force reduction of 70.2% with the structural optimization and 73.7% with the electrical driving optimization for the insertion operation into gelatin phantom, compared with a static needle.
  • Keywords
    acoustic impedance; acoustic resonance; acoustic signal processing; biological tissues; biomedical transducers; biomedical ultrasonics; biomimetics; needles; ultrasonic transducers; cancer biopsy; driving signal; electrical driving method; electrical driving optimization; gelatin phantom; geometrical dimensions; insertion operation; needle deflection; needle insertion; needle transducer structure; needle transducer vibration performance; penetration force; performance optimization; regional anaesthesia; resonant frequency; soft tissue; structural optimization; tissue mimics; transducer resonant impedance; transversal dimensions; ultrasonic needle actuating device; Acoustics; Force; Needles; Optimization; Resonant frequency; Transducers; Vibrations; medical ultrasonic needle transducer; needle intervention; numerical analysis; performance optimization;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2014 IEEE International
  • Conference_Location
    Chicago, IL
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2014.0202
  • Filename
    6931967