DocumentCode
1336456
Title
Fast Exact/Quasi-Exact FBP Algorithms for Triple-Source Helical Cone-Beam CT
Author
Lu, Yang ; Katsevich, Alexander ; Zhao, Jun ; Yu, Hengyong ; Wang, Ge
Author_Institution
Dept. of Biomed. Eng., Shanghai Jiao Tong Univ., Shanghai, China
Volume
29
Issue
3
fYear
2010
fDate
3/1/2010 12:00:00 AM
Firstpage
756
Lastpage
770
Abstract
Cardiac computed tomography (CT) has been improved over past years, but it still needs improvement for higher temporal resolution in the cases of high or irregular cardiac rates. Given successful applications of dual-source cardiac CT scanners, triple-source cone-beam CT seems a promising mode for cardiac CT. In this paper, we propose two filtered-backprojection algorithms for triple-source helical cone-beam CT. The first algorithm utilizes two families of filtering lines. These lines are parallel to the tangent of the scanning trajectory and the so-called L lines. The second algorithm utilizes two families of filtering lines tangent to the boundaries of the Zhao window and L lines, respectively, but it eliminates the filtering paths along the tangent of the scanning trajectory, thus reducing the required detector size greatly. The first algorithm is theoretically exact for r < 0.265 R and quasi-exact for 0.265 R ?? r < 0.495 R, and the second algorithm is quasi-exact for r < 0.495 R, where r and R denote the object radius and the trajectory radius, respectively. Both algorithms are computationally efficient. Numerical results are presented to verify and showcase the proposed algorithms.
Keywords
computerised tomography; filtering theory; medical image processing; L lines; cardiac computed tomography; fast exact FBP algorithms; filtered backprojection; quasiexact FBP algorithms; scanning trajectory; triple-source helical cone-beam CT; Biomedical engineering; Computed tomography; Data acquisition; Detectors; Filtering algorithms; Heart; Mathematics; Research and development; Sensor arrays; Spatial resolution; Cardiac CT; computed tomography (CT); cone-beam CT (CBCT); filtered backprojection (FBP); triple-source; Algorithms; Computer Simulation; Cone-Beam Computed Tomography; Image Processing, Computer-Assisted; Models, Theoretical; Phantoms, Imaging;
fLanguage
English
Journal_Title
Medical Imaging, IEEE Transactions on
Publisher
ieee
ISSN
0278-0062
Type
jour
DOI
10.1109/TMI.2009.2035617
Filename
5338022
Link To Document