Series-parallel versus true polyphase calibration of electricity meters
Author
Williams, R. Doug ; Arseneau, Rejean ; Belanger, J.M. ; Braun, Jim ; Carr, John M. ; Crampton, Stewart W. ; Domijan, Alex, Jr. ; Flessner, Jim ; Knight, R. Don ; Larson, Richard E. ; Meyer, Mark ; Millican, Herman ; McComb, T.R. ; Sisson, P.E.
Volume
13
Issue
2
fYear
1998
fDate
4/1/1998 12:00:00 AM
Firstpage
351
Lastpage
355
Abstract
This paper describes a study of single-phase-series-parallel versus polyphase calibration of electricity meters. Historically there has been some debate regarding the field performance of electromechanical electricity meters that have been calibrated using the series-parallel technique which is common in North America and other areas. A methodology was carefully established that would allow several meter testing laboratories to be verified and traceable to each other. Meters were tested and the data is presented along with data analysis and conclusions
Keywords
calibration; power meters; power system measurement; North America; electricity meters; electromechanical electricity meters; field performance; meter testing laboratories; polyphase calibration; single-phase-series-parallel calibration; true polyphase testing; Calibration; Coils; Electrical capacitance tomography; Instrumentation and measurement; Laboratories; Power industry; Power system measurements; Qualifications; Testing; Watthour meters;