DocumentCode
1339759
Title
Miniaturisation of a capillary electrophoresis microchip for the sensing of endocrine disruptors
Author
Ha, Keum-Won ; Joo, G.-S. ; Jha, Sumit Kumar ; Yeon, I.-J. ; Kim, Young-Sik
Author_Institution
Dept. of Nano Sci. & Eng., Myongji Univ., Yongin, South Korea
Volume
4
Issue
4
fYear
2010
fDate
12/1/2010 12:00:00 AM
Firstpage
103
Lastpage
108
Abstract
Although capillary electrophoresis amperometric detector (CE-AD) involving double-T microchannel configuration is a powerful analytical tool in terms of sensitivity and selectivity, its long microchannel configuration hinders further miniaturisation. Therefore a twisted CE microchannel configuration was used in the present study to fabricate CE-AD devices for detection of endocrine disruptors. The analyte separation time varied slightly for the twisted microchannel structure, whereas the detector sensitivities were similar for the two configurations. The conventional indium tin oxide amperometric detector in the device with twisted microchannel configuration was later modified with Prussian blue to enhance the sensitivity of detection.
Keywords
amperometric sensors; biomedical measurement; electrophoresis; indium compounds; microchannel flow; microfluidics; microsensors; ITO; analyte separation time; capillary electrophoresis amperometric detector; capillary electrophoresis microchip; double-T microchannel configuration; endocrine disruptor sensing; miniaturisation;
fLanguage
English
Journal_Title
Nanobiotechnology, IET
Publisher
iet
ISSN
1751-8741
Type
jour
DOI
10.1049/iet-nbt.2010.0006
Filename
5591551
Link To Document