    Partial self-calibration method for permittivity measurement using truncated coaxial cable

  • Author

    Migliore, M.D.

  • Author_Institution
    Dipt. di Ingegneria Elettronica e delle Telecommun., Naples Univ., Italy
  • Volume
    36
  • Issue
    15
  • fYear
    2000
  • fDate
    7/20/2000 12:00:00 AM
  • Firstpage
    1275
  • Lastpage
    1277
  • Abstract
    A calibration procedure is presented for measuring permittivity by means of a truncated coaxial cable. The procedure requires a short circuit, an open circuit into air and two liquids the permittivity of which need only be approximately known, given that the method is sound with respect to liquid standard uncertainties