  • Title

    KLT-based quality controlled compression of single-lead ECG

  • Author

    Blanchett, T. ; Kember, G.C. ; Fenton, G.A.

  • Author_Institution
    Dept. of Appl. Math., Tech. Univ. Nova Scotia, Halifax, NS, Canada
  • Volume
    45
  • Issue
    7
  • fYear
    1998
  • fDate
    7/1/1998 12:00:00 AM
  • Firstpage
    942
  • Lastpage
    945
  • Abstract
    An electrocardiogram (EGG) compression algorithm based on a combination of the Karhunen-Loeve transform (KLT) and multirate sampling is introduced. The use of multirate sampling reduces KLT computational times to those reported for wavelet-packet-based compression techniques. A beat-by-beat quality controlled compression criterion is shown to be necessary to ensure clinically adequate reconstruction of each beat. The resulting quality controlled algorithm efficiently achieves compression rates of approximately 30-40:1 for the MIT-BIH database.
  • Keywords
    electrocardiography; medical signal processing; quality control; transforms; Karhunen-Loeve transform-based quality controlled compression; beat-by-beat quality controlled compression criterion; clinically adequate reconstruction; compression rates; electrocardiogram compression algorithm; multirate sampling; quality controlled algorithm; single-lead ECG; Biomedical measurements; Databases; Eigenvalues and eigenfunctions; Electrocardiography; Error correction; Frequency; Karhunen-Loeve transforms; Mathematics; Sampling methods; Wavelet packets; Algorithms; Electrocardiography; Quality Control; Signal Processing, Computer-Assisted;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.686803
  • Filename
    686803