• DocumentCode
    1396646
  • Title

    Benefits of angular expression to reconstruction algorithms for collimators with spatially varying focal lengths

  • Author

    You, Jiangsheng ; Bao, Shanglian ; Liang, Zhengrong

  • Author_Institution
    Inst. of Heavy Ion Phys., Beijing Univ., China
  • Volume
    16
  • Issue
    5
  • fYear
    1997
  • Firstpage
    527
  • Lastpage
    531
  • Abstract
    Fan-beam collimators with spatially varying focal lengths are used in single photon emission computed tomography (SPECT) to improve the imaging sensitivity and to reduce the reconstruction artifacts resulting from the truncation of projection data. An angular representation of the detector coordinates for the projection data is adopted to investigate several aspects of the reconstruction problem for this type of collimation. A rebinning reconstruction algorithm is derived. We prove a conjecture posed by Zeng and Gullberg (1994) and obtain a simplified form of their backprojection filtering algorithm. Some computer simulations are presented to investigate the performance of the rebinning algorithm.
  • Keywords
    biomedical equipment; filtering theory; focusing; gamma-ray apparatus; gamma-ray detection; image reconstruction; medical image processing; single photon emission computed tomography; SPECT; angular expression; angular representation; backprojection filtering algorithm; collimators; computer simulations; detector coordinates; fan-beam collimators; imaging sensitivity; projection data truncation; rebinning reconstruction algorithm; reconstruction algorithms; reconstruction artifacts; single photon emission computed tomography; spatially varying focal lengths; Convolution; Detectors; Filtering algorithms; Geometry; Image reconstruction; Optical collimators; Physics; Radiology; Reconstruction algorithms; Single photon emission computed tomography; Algorithms; Artifacts; Computer Simulation; Humans; Image Enhancement; Image Processing, Computer-Assisted; Models, Biological; Phantoms, Imaging; Tomography, Emission-Computed, Single-Photon;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/42.640742
  • Filename
    640742