• DocumentCode
    1403498
  • Title

    Characteristic investigation of a static micro polymerase chain reaction chip based on in situ electrochemical detection

  • Author

    Tao Deng ; Mengwei Li ; Chenxu Zhao ; Jian Qin ; Cangran Guo ; Zewen Liu

  • Author_Institution
    Inst. of Microelectron., Tsinghua Univ., Beijing, China
  • Volume
    7
  • Issue
    12
  • fYear
    2012
  • fDate
    12/1/2012 12:00:00 AM
  • Firstpage
    1226
  • Lastpage
    1229
  • Abstract
    Aiming to meet the ever-increasing challenge of a low cost, portable and real-time sensitive detection polymerase chain reaction (PCR) system, a novel in situ electrochemical detection static micro PCR (EC-SμPCR) chip is designed and fabricated. The 10μ10μmm EC-SμPCR chip comprises 12 micro reaction chambers, and each 1500μ1000μ60μμm chamber consists of three modules: a micro heater, a micro temperature sensor and a micro three-electrode system (MTES). Heating, cooling and electrochemical cyclic voltammetry (CV) experiments were carried out to investigate the performance characteristics of each module of the chamber in the EC-SμPCR chip. The results showed that the heating rate of the chip was 1.05μCs-1 with only one micro heater, whereas the cooling rate was 1.01°Cs-1 with natural air cooling. It is worth noting that the bare MTES of the chamber directly detected the oxidation of single-stranded DNA (ssDNA, 10-6 M) with the CV method, furthermore, the oxidation peak current of ssDNA can be increased about 40 after a prepolarisation step. All three modules of the chamber in the EC-S PCR chip work efficiently and reliably, which reveals that the entire chip is ready for further analysis trials.
  • Keywords
    DNA; biochemistry; biosensors; cooling; electrochemical sensors; lab-on-a-chip; microfabrication; microsensors; molecular biophysics; oxidation; polymers; temperature sensors; voltammetry (chemical analysis); CV method; MTES; cooling; electrochemical cyclic voltammetry; heating rate; in situ electrochemical detection static micro PCR chip; micro temperature sensor; micro three-electrode system; microheater; oxidation; oxidation peak current; real-time sensitive detection polymerase chain reaction system; single-stranded DNA; static micro polymerase chain reaction chip;
  • fLanguage
    English
  • Journal_Title
    Micro & Nano Letters, IET
  • Publisher
    iet
  • ISSN
    1750-0443
  • Type

    jour

  • DOI
    10.1049/mnl.2012.0551
  • Filename
    6419600