• DocumentCode
    1413827
  • Title

    Performance Evaluation of a PEM Scanner Using the NEMA NU 4—2008 Small Animal PET Standards

  • Author

    Luo, Weidong ; Anashkin, Edward ; Matthews, Christopher G.

  • Author_Institution
    Naviscan Inc., San Diego, CA, USA
  • Volume
    57
  • Issue
    1
  • fYear
    2010
  • Firstpage
    94
  • Lastpage
    103
  • Abstract
    The recently published NEMA NU 4-2008 Standards has been specially designed for evaluating the performance of small animal PET scanners used in preclinical applications. In this paper, we report on the NU 4 performance of a clinical positron emission mammography (PEM) system. Since there are no PEM specific performance test protocols available, and the NU 2 protocol (intended for whole-body PET scanners) cannot be applied without modification due to the compact design of the PEM scanner, we decided to evaluate the NU 4 Standards as an alternative. We obtained the following results: Trans-axial spatial resolution 1.8 mm FWHM for high resolution reconstruction mode and 2.4 mm FWHM for standard resolution reconstruction mode with no significant variation within the field of view. The total system sensitivity was 0.16 cps/Bq. In image quality testing, the uniformity was found to be 3.9% STD at the standard resolution mode and 5.6% at the high resolution mode when measured with a 34 mm paddle separation. The NEMA NU 4-2008 Standards were found to be a practicable tool to evaluate the performance of the PEM scanner after some modifications to address the specifics of its detector configuration. Furthermore, the PEM scanner´s in-plane spatial resolution was comparable to other small animal PET scanners with good image quality.
  • Keywords
    image reconstruction; image resolution; mammography; medical image processing; positron emission tomography; NEMA NU 4-2008 small animal PET standards; NU 2 protocol; PEM scanner; clinical positron emission mammography system; distance 34 mm; high-resolution reconstruction mode; image quality testing; in-plane spatial resolution; paddle separation; standard resolution mode; transaxial spatial resolution; Animals; Image quality; Image reconstruction; Image resolution; Positron emission tomography; Protocols; Radioactive decay; Spatial resolution; Standards publication; Whole-body PET; NEMA performance standards; positron emission mammography; preclinical PET;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2009.2036847
  • Filename
    5409986