• DocumentCode
    1438572
  • Title

    Lynch-Ayers microwave dielectric loss bridge for use over a range of temperatures

  • Author

    Gelsthorpe, R.V. ; Bennett, R.G.

  • Author_Institution
    Chelsea College of Science and Technology, Department of Electronics, London, UK
  • Volume
    124
  • Issue
    6
  • fYear
    1977
  • fDate
    6/1/1977 12:00:00 AM
  • Firstpage
    572
  • Lastpage
    574
  • Abstract
    Precision measurements of losses in low-loss liquid dielectrics at microwave frequencies may be performed by means of a bridge method described by Lynch and Ayers. The paper describes modifications to such a bridge in order that measurements may be made over a range of temperatures. Results are presented for the temperature variation of loss angle at X-band in two very low-loss materials (cyclohexane and cycloheptane). The errors arising in the measurements are analysed; the overall likely error in the determination being ±(1¿rad or 2.4%).
  • Keywords
    bridge instruments; dielectric loss measurement; measurement errors; microwave measurement; Lynch Ayers microwave bridge; X-band; dielectric loss measurement; error; liquid dielectrics; loss angle; modifications; precision;
  • fLanguage
    English
  • Journal_Title
    Electrical Engineers, Proceedings of the Institution of
  • Publisher
    iet
  • ISSN
    0020-3270
  • Type

    jour

  • DOI
    10.1049/piee.1977.0120
  • Filename
    5252784