• DocumentCode
    1464786
  • Title

    Development of a 3.5-mm Coaxial Microcalorimeter for RF and Microwave Power Standards at KRISS

  • Author

    Kwon, Jae-Yong ; Kang, Tae-Weon ; Kim, Jeong-Hwan ; Kang, Jin-Seob

  • Author_Institution
    Center for Electromagn. Wave, Korea Res. Inst. of Stand. & Sci., Daejeon, South Korea
  • Volume
    60
  • Issue
    7
  • fYear
    2011
  • fDate
    7/1/2011 12:00:00 AM
  • Firstpage
    2609
  • Lastpage
    2614
  • Abstract
    We have developed a 3.5-mm coaxial microcalorimeter that will become a national power standard up to 26.5 GHz. Several core parts such as a thermostat, an adiabatic line, a thermopile, and reference standards are described in detail. Some measurement data are shown for a system performance check. A finite-element method simulation of the adiabatic line is shown. A model of the effect of the adiabatic line on thermopile measurement is introduced. Effective efficiency at selected frequencies is presented for the reference standards.
  • Keywords
    calorimeters; finite element analysis; power measurement; thermopiles; thermostats; RF power; adiabatic line; coaxial microcalorimeter; finite-element method simulation; microwave power; reference standards; size 3.5 mm; thermopile measurement; thermostat; Electromagnetics; Power measurement; Radio frequency; Standards; Temperature control; Temperature measurement; Thermostats; Calibration; coaxial transmission lines; power measurement; temperature control; thermistor; thermoelectric devices;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2011.2113850
  • Filename
    5723747