DocumentCode
1469229
Title
An RFID-Based Closed-Loop Wireless Power Transmission System for Biomedical Applications
Author
Kiani, Mehdi ; Ghovanloo, Maysam
Author_Institution
GT-Bionics Lab., Georgia Inst. of Technol., Atlanta, GA, USA
Volume
57
Issue
4
fYear
2010
fDate
4/1/2010 12:00:00 AM
Firstpage
260
Lastpage
264
Abstract
This brief presents a standalone closed-loop wireless power transmission system that is built around a commercial off-the-shelf (COTS) radio-frequency identification (RFID) reader (TRF7960) operating at 13.56 MHz. It can be used for inductively powering implantable biomedical devices in a closed loop. Any changes in the distance and misalignment between transmitter and receiver coils in near-field wireless power transmission can cause a significant change in the received power, which can cause either a malfunction or excessive heat dissipation. RFID circuits are often used in an open loop. However, their back telemetry capability can be utilized to stabilize the received voltage on the implant. Our measurements showed that the delivered power to the transponder was maintained at 11.2 mW over a range of 0.5 to 2 cm, while the transmitter power consumption changed from 78 mW to 1.1 W. The closed-loop system can also oppose voltage variations as a result of sudden changes in the load current.
Keywords
biomedical equipment; closed loop systems; inductive power transmission; radiofrequency identification; radiofrequency integrated circuits; telemedicine; RFID circuits; RFID-based closed-loop wireless power transmission system; TRF7960; back telemetry; biomedical applications; frequency 13.56 MHz; implantable biomedical devices; near-field wireless power transmission; radio-frequency identification reader; Closed-loop inductive power transmission; implantable biomedical devices; radio-frequency identification (RFID) reader;
fLanguage
English
Journal_Title
Circuits and Systems II: Express Briefs, IEEE Transactions on
Publisher
ieee
ISSN
1549-7747
Type
jour
DOI
10.1109/TCSII.2010.2043470
Filename
5446413
Link To Document