• DocumentCode
    1476182
  • Title

    Visible-LED pumped fiber-optic temperature sensor

  • Author

    Augousti, Andreas T. ; Grattan, Kenneth T V ; Palmer, Andrew W.

  • Author_Institution
    Sch. of Electr. Eng. & Appl. Phys., City Univ., London, UK
  • Volume
    37
  • Issue
    3
  • fYear
    1988
  • fDate
    9/1/1988 12:00:00 AM
  • Firstpage
    470
  • Lastpage
    472
  • Abstract
    The development and calibration of a visible LED-pumped, photoluminescent optical-fiber sensor to measure temperature using the crystal alexandrite is reported. Utilizing the dependence of the fluorescent decay time on temperature, this instrument is accurate to ±1°C over a range 20-150°C, with a response time of 1 s
  • Keywords
    calibration; fibre optic sensors; light emitting diodes; temperature measurement; 1 s; 20 to 150 degC; calibration; crystal alexandrite; fluorescent decay time; photoluminescent optical-fiber sensor; temperature sensor; visible LED pumped fibre optic sensor; Calibration; Fluorescence; Instruments; Optical fiber sensors; Optical pumping; Optical sensors; Temperature dependence; Temperature distribution; Temperature measurement; Temperature sensors;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.7480
  • Filename
    7480