Abstract
A description is given of the neutron counter system used in a search for neutrons at high-sensitivity, the negative results obtained, and the identification of some potentially misleading artifacts of neutron detection. More than 30 different test cell configurations were tested to cover the full range of the pertinent experimental variables. The design and characteristics of the neutron counter are discussed. The significant features of this neutron search technique are: (1) three independent neutron detectors of relatively high efficiency were used to simultaneously monitor the test cells; (2) the detectors were calibrated using a 252Cf neutron source in place of the test cells; and (3) the experiments were performed in an underground laboratory to reduce the neutron background induced by cosmic rays. The overall efficiency of the counter for 252Cf neutrons was 9.2 to 15%, depending on the experimental arrangement. No evidence was observed for significant neutron emission from any of the more than 30 test cell configurations that were monitored for a total of 339 h in the underground laboratory
Keywords
calibration; neutron detection and measurement; particle detectors; 339 h; 252Cf neutron source; calibration; cold fusion; design; efficiency; neutron background; neutron counter system; Counting circuits; Detectors; Deuterium; Laboratories; Monitoring; Neutrons; Polyethylene; Radioactive decay; Temperature; Testing;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.57360
Filename
57360
Link To Document