Title
A noise-protected digital heart ratemeter
Author
Twizell, R.J.
Author_Institution
D.S.I.R., Auckland Industrial Development Division, Auckland, New Zealand
Volume
45
Issue
4
fYear
1975
fDate
4/1/1975 12:00:00 AM
Firstpage
155
Lastpage
160
Abstract
In many experimental and diagnostic studies of heart rate, a simple averaging ratemeter is not satisfactory. This is due to the difficulty of obtaining an instrument response sufficiently fast to follow the rapid changes in rate, and at the same time keeping the ripple to an acceptable level. This paper describes an instantaneous-frequency cardiotachometer with some unique design features. It focuses particular attention on the digital period-to-frequency converter and a protection system designed to reject spurious noise pulses, such as those caused by muscle artefacts during exercise studies and those occurring during foetal heart rate monitoring.
Keywords
biomedical electronics; cardiology; digital instrumentation; frequency measurement; patient diagnosis; tachometers; cardiotachometer; diagnostic; digital heart ratemeter; noise protection;
fLanguage
English
Journal_Title
Radio and Electronic Engineer
Publisher
iet
ISSN
0033-7722
Type
jour
DOI
10.1049/ree.1975.0029