DocumentCode
1480487
Title
Monte Carlo and experimental evaluation of BC-454 for use as a multienergy neutron detector
Author
McElhaney, S.A. ; Bauer, M.L. ; Chiles, M.M. ; Emmett, M.B. ; Todd, J.H.
Author_Institution
Oak Ridge Nat. Lab., TN, USA
Volume
37
Issue
3
fYear
1990
fDate
6/1/1990 12:00:00 AM
Firstpage
1351
Lastpage
1358
Abstract
A boron-substituted plastic scintillator has been evaluated for use as a multienergy neutron detector in the Navy RADIAC Development Program. Monte Carlo codes were run for a 1 in. diameter×1 in. long scintillator and a 2 in. diameter×2 in. long scintillator. Additional calculations were run with each scintillator encased in 2 in. of polyethylene moderator. Response curves for neutron sensitivity as a function of energy were plotted from these calculations. Results show a close approximation between the analytically modeled response and the response obtained from a commercially available multienergy neutron detector. The scintillators also were evaluated experimentally for comparison with Monte Carlo results. Each scintillator was coupled to a photomultiplier tube and subjected to neutron irradiation. Results indicate that the experimental configuration used was rejecting proton recoil events <1 MeV. This proton recoil cutoff energy resulted in an underrepresented intermediate energy region
Keywords
Monte Carlo methods; neutron detection and measurement; scintillation counters; 1 MeV; 1 in; 2 in; BC-454; Monte Carlo; intermediate energy; multienergy neutron detector; neutron sensitivity; photomultiplier tube; plastic scintillator; polyethylene moderator; proton recoil events; response curves; Absorption; Boron; Detectors; Electrons; Laboratories; Linear accelerators; Monte Carlo methods; Neutrons; Plastics; Protons;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.57387
Filename
57387
Link To Document